Elastodontic Devices in Orthodontics: An In-Vitro Study on Mechanical Deformation under Loading
Articolo
Data di Pubblicazione:
2022
Citazione:
Elastodontic Devices in Orthodontics: An In-Vitro Study on Mechanical Deformation under Loading / Quinzi, Vincenzo; Gallusi, Gianni; Carli, Elisabetta; Pepe, Francesca; Rastelli, Elena; Tecco, Simona. - In: BIOENGINEERING. - ISSN 2306-5354. - 9:7(2022), p. 282. [10.3390/bioengineering9070282]
Abstract:
The purpose of the present study was to evaluate the mechanical resistance of elastodontic devices (ED): their maximum compression loads and plastic deformation under loading (percentage). An Instron universal machine (Model 3365, Instron, Industrial Product Group, Grove City, PA, USA) was employed with a 100 N load cell and with Bluehill software for loading analyses. Each device was submitted to a five-cycles test. The following ED were evaluated: A.M.C.O.P. (Micerium, Genova, Italy) in red color, in orange color, and in blue color; HealthyStart (Ortho-Tain, Winnetka, IL, USA), and T4K (TM) phase 1 (Myofunctional Research Co., Helensvale, Australia). During the five-cycles test, the Ortho-Tain device delivered the greatest compression load (7.56 N), with the lowest percentage of deformation (0.95%). For all devices, a slight plastic deformation of the material was registered, ranging from 0.95% to 1.75%. For the T4K device it was not possible to complete the five-cycles test. For all the analyzed ED, a slight plastic deformation under loading was registered, that in all cases can be considered clinically acceptable. Further studies are needed to test the appliances after clinical usage.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
compression load; deformation; elastodontic devices; mechanical properties; pre-formed appliances; preventive orthodontics; removable orthodontic appliances
Elenco autori:
Quinzi, Vincenzo; Gallusi, Gianni; Carli, Elisabetta; Pepe, Francesca; Rastelli, Elena; Tecco, Simona
Link alla scheda completa:
Pubblicato in: