Skip to Main Content (Press Enter)

Logo UNISR
  • ×
  • Home
  • Persone
  • Pubblicazioni
  • Facoltà
  • Ambiti Di Ricerca

UNIFIND
Logo UNISR

|

UNIFIND

unisr.it
  • ×
  • Home
  • Persone
  • Pubblicazioni
  • Facoltà
  • Ambiti Di Ricerca
  1. Pubblicazioni

Outcomes from a recovery protocol for patients with continuous-flow left ventricular assist devices

Articolo
Data di Pubblicazione:
2019
Citazione:
Outcomes from a recovery protocol for patients with continuous-flow left ventricular assist devices / Knierim, J.; Heck, R.; Pieri, M.; Schoenrath, F.; Soltani, S.; Stawowy, P.; Dreysse, S.; Stein, J.; Muller, M.; Mulzer, J.; Dandel, M.; Falk, V.; Krabatsch, T.; Potapov, E.. - In: THE JOURNAL OF HEART AND LUNG TRANSPLANTATION. - ISSN 1053-2498. - 38:4(2019), pp. 440-448. [10.1016/j.healun.2018.11.001]
Abstract:
BACKGROUND: In this retrospective analysis we evaluated a standardized echocardiographic assessment and an invasive technique for patient selection for successful continuous-flow left ventricular assist device (CF-LVAD) explantation. METHODS: Inclusion criteria for LVAD recovery assessment were: clinically stable condition; LVAD support for >6 months; physical activity; normal echocardiography findings; and no more than mild valvular disease and aortic valve opening. In a second step, echocardiography was performed under CF-LVAD reduction and stop conditions (PStopE). In the third step, patients who presented with stable parameters underwent right heart catheterization under CF-LVAD stoppage and occlusion of the outflow graft with a balloon catheter. Criteria for explantation were normal pulmonary artery pressure and pulmonary capillary wedge pressure <16 mmHg. RESULTS: Thirty-three of 424 patients entered the second step of evaluation and 20 entered the third step. Fourteen presented positive results and the pump was successfully explanted. The PCWP at baseline was 8.5 (2.8) mmHg in the explantation group and 10.6 (2.8) mmHg in the non-explantation group (p = 0.105). It increased to 10.9 (3.0) mmHg vs 20.8 (4.9) mmHg under outflow graft occlusion. The wedge pressure was significantly higher in the non-explantation group (p < 0.001). Median duration of follow-up after explantation was 9.74 (interquartile range 4.3 to 20.60) months, with survival of 93%. CONCLUSIONS: The protocol presented is feasible and safe. The criteria applied provide good patient selection for sustained mid-term myocardial recovery after LVAD explantation.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
explantation; graft occlusion; left ventricular assist device; pump stop; recovery; right heart catheterization; weaning
Elenco autori:
Knierim, J.; Heck, R.; Pieri, M.; Schoenrath, F.; Soltani, S.; Stawowy, P.; Dreysse, S.; Stein, J.; Muller, M.; Mulzer, J.; Dandel, M.; Falk, V.; Krabatsch, T.; Potapov, E.
Autori di Ateneo:
PIERI MARINA LAURA GRAZIA
Link alla scheda completa:
https://iris.unisr.it/handle/20.500.11768/179337
Pubblicato in:
THE JOURNAL OF HEART AND LUNG TRANSPLANTATION
Journal
  • Utilizzo dei cookie

Realizzato con VIVO | Designed by Cineca | 26.5.1.0