Comparative Evaluation of the Bruker Biotyper and Vitek MS Matrix-Assisted Laser Desorption Ionization–Time Of Flight (MALDI-TOF) Mass Spectrometry Systems for Identification of Yeasts of Medical Importance

Academic Article
Publication Date:
2013
abstract:
We report the first comparative evaluation between the Bruker Biotyper MS (BMS) and the Vitek MS (VMS) for the identification of yeasts. Correct identification rate at the species level was comparable using the commercial databases (89.8% vs. 84.3% p=0.712), but higher for BMS using an in-house-extended database (100% vs. 84.3% p=0.245). Importantly, the rate of misidentification was significantly higher for VMS (1% vs. 12.1% p<0.0001), including the rate of major errors (0% vs. 4.5% p=0.0036).
Iris type:
1.1 Articolo in rivista
List of contributors:
Mancini, Nicasio; De Carolis, E; Infurnari, L; Vella, A; Clementi, Nicola; Vaccaro, L; Ruggeri, A; Posteraro, B; Burioni, Roberto; Clementi, Massimo; Sanguinetti, M.
Authors of the University:
BURIONI ROBERTO
CLEMENTI NICOLA
CLEMENTI MASSIMO
Handle:
https://iris.unisr.it/handle/20.500.11768/47583
Published in:
JOURNAL OF CLINICAL MICROBIOLOGY