Angulation and curvature of aortic landing zone affect implantation depth in transcatheter aortic valve implantation

Articolo
Data di Pubblicazione:
2024
Citazione:
Angulation and curvature of aortic landing zone affect implantation depth in transcatheter aortic valve implantation / Gorla, R.; Oliva, O. A.; Arzuffi, L.; Milani, V.; Saitta, S.; Squillace, M.; Poletti, E.; Tusa, M.; Votta, E.; Brambilla, N.; Testa, L.; Bedogni, F.; Sturla, F.. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - 14:1(2024). [10.1038/s41598-024-61084-5]
Abstract:
In transcatheter aortic valve implantation (TAVI), final device position may be affected by device interaction with the whole aortic landing zone (LZ) extending to ascending aorta. We investigated the impact of aortic LZ curvature and angulation on TAVI implantation depth, comparing short-frame balloon-expanding (BE) and long-frame self-expanding (SE) devices. Patients (n = 202) treated with BE or SE devices were matched based on one-to-one propensity score. Primary endpoint was the mismatch between the intended (HPre) and the final (HPost) implantation depth. LZ curvature and angulation were calculated based on the aortic centerline trajectory available from pre-TAVI computed tomography. Total LZ curvature (kLZ,tot) and LZ angulation distal to aortic annulus (αLZ,Distal) were greater in the SE compared to the BE group (P < 0.001 for both). In the BE group, HPost was significantly higher than HPre at both cusps (P < 0.001). In the SE group, HPost was significantly deeper than HPre only at the left coronary cusp (P = 0.013). At multivariate analysis, αLZ,Distal was the only independent predictor (OR = 1.11, P = 0.002) of deeper final implantation depth with a cut-off value of 17.8°. Aortic LZ curvature and angulation significantly affected final TAVI implantation depth, especially in high stent-frame SE devices reporting, upon complete release, deeper implantation depth with respect to the intended one.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
Aortic stenosis; Innovation; Risk stratification; TAVI
Elenco autori:
Gorla, R.; Oliva, O. A.; Arzuffi, L.; Milani, V.; Saitta, S.; Squillace, M.; Poletti, E.; Tusa, M.; Votta, E.; Brambilla, N.; Testa, L.; Bedogni, F.; Sturla, F.