Technical Tips for the Implantation of the Gore® Excluder® Abdominal Aortic Aneurysm and Iliac Branch Endoprosthesis in Short Aortoiliac Anatomies
Academic Article
Publication Date:
2019
abstract:
The purpose of this article is to describe the techniques used to deploy a Gore® Excluder® abdominal aortic aneurysm (AAA) and iliac branch endoprosthesis (IBE) in a short aortoiliac anatomy, expanding treatment options of a patient otherwise unfit for endovascular repair. Four strategies were used to shorten the total length of the components, achieving a reduction of over 3 cm during deployment: a pull-pull technique on a through-and-through guidewire to accommodate the IBE onto the iliac bifurcation, deployment in a “cross-limb” configuration, foreshortening of the iliac bridging component during release, and final proximal shortening of the AAA main body thanks to the C3 repositionable delivery system.
Iris type:
1.1 Articolo in rivista
Keywords:
Aged, 80 and over; Aorta, Abdominal; Aortic Aneurysm, Abdominal; Blood Vessel Prosthesis Implantation; Endovascular Procedures; Humans; Iliac Aneurysm; Iliac Artery; Male; Prosthesis Design; Treatment Outcome; Blood Vessel Prosthesis
List of contributors:
Fernandez Prendes, C.; Mascia, D.; Melissano, G.; Chiesa, R.; Kahlberg, A.
Published in: