MiniMed 780G Six-Month Use in Children and Adolescents with Type 1 Diabetes: Clinical Targets and Predictors of Optimal Glucose Control

Articolo
Data di Pubblicazione:
2023
Citazione:
MiniMed 780G Six-Month Use in Children and Adolescents with Type 1 Diabetes: Clinical Targets and Predictors of Optimal Glucose Control / Lombardo, F.; Passanisi, S.; Alibrandi, A.; Bombaci, B.; Bonfanti, R.; Delvecchio, M.; Di Candia, F.; Mozzillo, E.; Piccinno, E.; Piona, C. A.; Rigamonti, A.; Scialabba, F.; Maffeis, C.; Salzano, G.. - In: DIABETES TECHNOLOGY & THERAPEUTICS. - ISSN 1520-9156. - 25:6(2023), pp. 404-413. [10.1089/dia.2022.0491]
Abstract:
Background: The aim of this multicenter observational real-world study was to investigate glycemic outcomes in children and adolescents with type 1 diabetes over the first 6-month use of MiniMed™ 780G. The secondary objective was to evaluate demographic and clinical factors that may be significantly associated with the achievement of therapeutic goals. Methods: Demographic, anamnestic, and clinical data of study participants were collected at the time of enrollment. Data on ambulatory glucose profile were acquired at 3 and 6 months after activating automatic mode. Aggregated glucose metrics and device settings of the entire study period were analyzed to identify predictors of optimal glycemic control, assessed by the concomitant achievement of time in range (TIR) >70%, coefficient of variation (CV) <36%, glucose management indicator (GMI) <7%, and time below range (TBR) <4%. Results: Our study cohort consisted of 111 children and adolescents (54.1% female) aged 7-18 years. All the most relevant clinical targets were achieved according to recommendations from the International Consensus both at 3 and 6 months. When considering aggregated data, primary goals in terms of TIR, CV, GMI, and TBR were achieved, respectively, by 72.1%, 74.8%, 68.5%, and 74.8% of participants. In addition, 44 individuals (39.6%) concomitantly addressed all the above clinical targets. Regression analysis revealed that older age, briefer duration of disease, and shorter active insulin time were significant predictors of optimal glucose control. Comparing two groups of individuals stratified according to the glycated hemoglobin (HbA1c) mean value in the year preceding MiniMed 780G use, achieving glycemic targets was observed in the subgroup with lower HbA1c. Conclusions: Our study highlights the effectiveness and safety of MiniMed 780G in the pediatric population. More extensive and personalized training on advanced hybrid closed-loop use should be considered for younger people and those with long disease duration.
Tipologia CRIS:
1.1 Articolo in rivista
Elenco autori:
Lombardo, F.; Passanisi, S.; Alibrandi, A.; Bombaci, B.; Bonfanti, R.; Delvecchio, M.; Di Candia, F.; Mozzillo, E.; Piccinno, E.; Piona, C. A.; Rigamonti, A.; Scialabba, F.; Maffeis, C.; Salzano, G.
Autori di Ateneo:
BONFANTI RICCARDO
Link alla scheda completa:
https://iris.unisr.it/handle/20.500.11768/170357
Pubblicato in:
DIABETES TECHNOLOGY & THERAPEUTICS
Journal