Leveraging the potential of 1.0-mm i.d. columns in UHPLC-HRMS-based untargeted metabolomics

Articolo
Data di Pubblicazione:
2025
Citazione:
Leveraging the potential of 1.0-mm i.d. columns in UHPLC-HRMS-based untargeted metabolomics / La Gioia, D., Salviati, E., Basilicata, M.G., Felici, C., Botrugno, O.A., Tonon, G., Sommella, E., Campiglia, P.. - In: ANALYTICAL AND BIOANALYTICAL CHEMISTRY. - ISSN 1618-2642. - 417:13(2025), pp. 2837-2847. [10.1007/s00216-024-05588-z]
Abstract:
Untargeted metabolomics UHPLC-HRMS workflows typically employ narrowbore 2.1-mm inner diameter (i.d.) columns. However, the wide concentration range of the metabolome and the need to often analyze small sample amounts poses challenges to these approaches. Reducing the column diameter could be a potential solution. Herein, we evaluated the performance of a microbore 1.0-mm i.d. setup compared to the 2.1-mm i.d. benchmark for untargeted metabolomics. The 1.0-mm i.d. setup was implemented on a micro-UHPLC system, while the 2.1-mm i.d. on a standard UHPLC, both coupled to quadrupole-orbitrap HRMS. On polar standard metabolites, a sensitivity gain with an average 3.8-fold increase over the 2.1-mm i.d., along with lower LOD (LODavg 1.48 ng/mL vs. 6.18 ng/mL) and LOQ (LOQavg 4.94 ng/mL vs. 20.60 ng/mL), was observed. The microbore method detected and quantified all metabolites at LLOQ with respect to 2.1, also demonstrating good repeatability with lower CV% for retention times (0.29% vs. 0.63%) and peak areas (4.65% vs. 7.27%). The analysis of various samples, in both RP and HILIC modes, including different plasma volumes, dried blood spots (DBS), and colorectal cancer (CRC) patient-derived organoids (PDOs), in full scan-data dependent mode (FS-DDA) reported a significant increase in MS1 and MS2 features, as well as MS/MS spectral matches by 38.95%, 39.26%, and 18.23%, respectively. These findings demonstrate that 1.0-mm i.d. columns in UHPLC-HRMS could be a potential strategy to enhance coverage for low-amount samples while maintaining the same analytical throughput and robustness of 2.1-mm i.d. formats, with reduced solvent consumption.
Tipologia CRIS:
1.1 Articolo in rivista
Elenco autori:
La Gioia, D.; Salviati, E.; Basilicata, M. G.; Felici, C.; Botrugno, O. A.; Tonon, G.; Sommella, E.; Campiglia, P.
Autori di Ateneo:
TONON GIOVANNI
Link alla scheda completa:
https://iris.unisr.it/handle/20.500.11768/185963
Link al Full Text:
https://iris.unisr.it//retrieve/handle/20.500.11768/185963/330033/unpaywall-bitstream--18218404.pdf
Pubblicato in:
ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Journal
  • Dati Generali

Dati Generali

URL

https://link.springer.com/article/10.1007/s00216-024-05588-z
  • Utilizzo dei cookie

Realizzato con VIVO | Designed by Cineca | 26.7.0.0